COVIDIEN LP - SUPERDIMENSION INC SUPERDIMENSION INREACH SYSTEM; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
|
Back to Search Results |
|
Model Number AAS00161-20U |
Device Problem
Programming Issue (3014)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/07/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
A component of the system, software data, has been requested but has not yet been received.A review of the appropriate device history records of the superdimension console indicates this device was released meeting all quality specifications.Out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, when they tried to plan or to input registration points there was planning issue and the 3d map was not generated.They were unable to complete the super dimension portion of the case.It is unknown if the patient was under general anesthesia.There was no report of patient injury.
|
|
Manufacturer Narrative
|
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Device component, planning files, and the patient ct scan were returned for analysis.The ct scan was loaded, 3d tree was generated and adding registration points was enabled and no software issue was found.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|