| Model Number |
M0068404010 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation results: one accumax 200 laser fiber was received for evaluation.Visual examination revealed that the exposed glass fiber tip measured approximately 3.5 mm and appeared unused.Additional examination under magnification confirmed that the tip was unused.There was no damage along the body of the fiber.Visual examination under magnification revealed that there was debris on the fiber face within the sma connector and only a small amount of light was coming through.The condition of the returned device would cause insufficient energy transmission during ablation thereby confirming the event.The noted damages indicate difficulty was experienced during the procedure and are likely due to anatomical or procedural factors such as tortuous anatomy or maneuvering of the device.Therefore, a review and analysis of all available information indicated that the most probable root cause is operational context.A review of the device history record (dhr) was performed; no anomalies were noted.A search of the complaint database revealed that no similar complaints exist for the specified lot.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that an accumax 200 laser fiber was used during a stone lithotripsy procedure performed on (b)(6) 2018.According to the complainant, during the procedure and outside the patient, the nurse noted that there was no laser energy coming out from the fiber when connected to the laser console.The procedure was completed with another accumax 200 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.This event has been deemed an mdr-reportable event based on investigation results which revealed that there was debris on the fiber face within the sma connector.
|
| |
|
Search Alerts/Recalls
|