| Model Number |
72201213 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the knot manipulators cut the ultrabraid #2 suture during the knotting.No significant delay or patient injury were reported.A back up was available.
|
| |
|
Additional Manufacturer Narrative
|
|
Examination was not possible, as the device has not been returned.The investigation could not draw any conclusions about the reported event without the return of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
One knot manipulator was returned for evaluation.Visual assessment of the device showed the distal tip to be smooth, no sharp edges.The device was tested using two strands of #2 braided suture.The device did not damage the suture in any manner.
|
| |
|