| Model Number |
D284TRK |
| Medical Device Problem Codes |
Mechanical Problem (1384); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 4196 lead implanted (b)(6) 2010.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited irregular and inconsistent impedance on both defibrillation coils as seen through the cardiac resynchronization therapy defibrillator (crt-d), possibly due to impending fracture.The crt-d was explanted and replaced.Multiple impedance measurements were taken through the new device and were within normal limits.The physician requested confirmation of a header issue with the replaced device.It was also reported that rising impedance was observed on the right atrial (ra) lead.No action was taken and the lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.If information is provided in the future, a supplemental report will be issued.
|
| |
|