| Catalog Number |
0066815 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The immucor laboratory tested a returned blood sample from the customer site with retention product on 01jun2018 which duplicated the customer's unexpectedly negative findings.The blood sample had a total protein of 8.6 g/dl, alb of 2.8 g/dl and globulin of 5.8 g/dl.The immucor internal identification number for this report is mdr652100.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a (b)(6) customer site reported an unexpectedly negative antibody screen and antibody panel, when using capture-r ready-screen (3) and capture-r ready-id with a galileo echo 2.0 instrument.
|
| |
|