| Model Number |
48130100 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while turning the rod, the rotation key and forceps broke.The surgery was completed successfully with alternate tools in the surgery room.No adverse consequences to the patient were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that while turning the rod, the rotation key and forceps broke.The surgery was completed successfully with alternate tools in the surgery room.No adverse consequences to the patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial reporter was updated from physician/surgeon to clinical support employee.
|
| |
|
Event or Problem Description
|
|
It was reported that while turning the rod, the rotation key and forceps broke.The surgery was completed successfully with alternate tools in the surgery room.No adverse consequences to the patient were reported.
|
| |
|
Search Alerts/Recalls
|