Catalog Number STT-GF-001
Device Problems Use of Device Problem (1670); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that the display device showed a transmitter failed error on (b)(6) 2018.The device has been received for investigation.A follow-up will be submitted upon completion of device investigation.No additional patient or event information is available.
 
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that the display device showed a transmitter failed error on (b)(6) 2018.The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Voltage testing was performed, and the test failed.A review of the downloaded data log did not confirm the reported event of failed transmitter error.A probable cause could not be determined.No additional patient or event information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7643072
MDR Text Key112720306
Report Number3004753838-2018-070142
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberSTT-GF-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/23/2018
Initial Date FDA Received06/27/2018
Supplement Dates Manufacturer Received08/27/2018
Supplement Dates FDA Received09/10/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age20 YR
-
-