| Catalog Number |
1012452-08 |
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that during device unpacking and prior to use, the 3.75 x 8 mm nc trek plastic holder was noted to be kinked.While attempting to remove the device from the holder, the hub detached from the catheter shaft.There was no damage noted to the outer chipboard box.There was no patient involvement.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual inspection was performed on the returned device.The reported kinked hoop/coil and the separation were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported kinked hoop/coil and the separation appear to be related to a potential product quality issue.Continued assessment of this issue per site operating procedures is being performed.These devices will continue to be monitored.
|
| |
|