• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (PRINGY) JUVEDERM ULTRA XC TSK US; IMPLANT, DERMAL, FOR AESTHETIC USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (PRINGY) JUVEDERM ULTRA XC TSK US; IMPLANT, DERMAL, FOR AESTHETIC USE Back to Search Results
Catalog Number 94154
Device Problems Detachment Of Device Component (1104); Expulsion (2933)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/07/2018
Event Type  malfunction  
Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Device labeling: precautions: ¿failure to comply with the needle attachment instructions could result in needle disengagement and/or product leakage at the luer-lok® and needle hub connection.Instructions for use: ¿if the needle is blocked, do not increase the pressure on the plunger rod.Instead, stop the injection and replace the needle.How supplied: ¿juvéderm® ultra xc injectable gel is supplied in individual treatment syringes with 30 g needles for single-patient use and ready for injection (implantation).The volume in each syringe is as stated on the syringe label and on the carton.The contents of the syringe are sterile and non-pyrogenic.Do not resterilize.Do not use if package is opened or damaged.
 
Event Description
Healthcare professional reported a syringe of juvéderm® ultra xc had the needle pop out and the gel went on the patient's face.Patient contact did occur.No injuries were reported.The packaged needle was used.
 
Manufacturer Narrative
Additional information: device analysis summary: visual analysis of the device indicates no defect observed.
 
Event Description
Healthcare professional reported a syringe of juvéderm® ultra xc had the needle pop out and the gel went on the patient's face.Patient contact did occur.No injuries were reported.The packaged needle was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUVEDERM ULTRA XC TSK US
Type of Device
IMPLANT, DERMAL, FOR AESTHETIC USE
Manufacturer (Section D)
ALLERGAN (PRINGY)
route de promery
zone artisanale de pre-mairy
pringy 74370
FR  74370
Manufacturer (Section G)
ALLERGAN (PRINGY)
route de promery
zone artisanale de pre-mairy
pringy 74370
FR   74370
Manufacturer Contact
suzanne wojcik
301 w howard lane
suite 100
austin, TX 78753
7372473605
MDR Report Key7644277
MDR Text Key112566436
Report Number3005113652-2018-00791
Device Sequence Number1
Product Code LMH
UDI-Device Identifier30888628000081
UDI-Public30888628000081
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/20/2019
Device Catalogue Number94154
Device Lot NumberH24LA70666
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age36 YR
-
-