Model Number MMT-332A
Device Problem Bent (1059)
Patient Problem Hyperglycemia (1905)
Event Date 04/16/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
Customer reported via phone call that they experienced high blood glucose level of 400 mg/dl.The customer had symptoms such as ketones and treated with manual injection.Customer's blood glucose value was 492 mg/dl at the time of the call.Customer reported that the high blood glucose levels began on (b)(6) 2018.Customer declined to troubleshoot for high readings.Troubleshooting was performed.The drive support cap appeared normal.There were no leaks or air bubbles.There were no site or insulin issues.Customer reported that the infusion set cannula was bent and there were changes in the device settings.The insulin pump passed the displacement test.Customer was advised to change the infusion set, reservoir and insulin and to treat per healthcare professional's recommendation.The product was not returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESERVOIR 3ML MMT-332A
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7644631
MDR Text Key112544512
Report Number3004209178-2018-85915
Device Sequence Number1
Product Code FRN
Combination Product (y/n)N
Reporter Country CodeCO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-332A
Device Catalogue NumberMMT-332A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-