JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS CL; MONOFOCAL IOLS Back to Search Results
Model Number Z9002
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the tecnis monofocal intraocular lens (model z9002, +18.0 diopter) was implanted and then the surgeon realized that the posterior capsule would not support the lens hence it was removed and replaced with an anterior chamber lens (non-johnson&johnson lens).Both the haptics were detached during the removal process.Additional information was received and it was learnt that during the removal of the lens, the incision was enlarged.No capsule tear and no other medical/surgical intervention was performed.The patient was reported doing good when discharged.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes, returned to manufacturer on: 07/20/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The lens was observed under microscope, and only half of the lens was returned to inspect, and the half that we receive had the haptic bent.The lens optic was observed cut in half and with residue.The complaint issue reported as haptic detached was not verified however due the condition of the lens received the complaint could not be related to the manufacturing process it could be related to the handling process; therefore, a product quality deficiency was not identified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7644797
Report Number2648035-2018-00916
Device Sequence Number275109
Product Code HQL
UDI-Device Identifier05050474530652
UDI-Public(01)05050474530652(17)220718
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/24/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date07/18/2022
Device Model NumberZ9002
Device Catalogue NumberZ900200180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 05/31/2018
Supplement Date Received by Manufacturer08/03/2018
Initial Report FDA Received Date06/28/2018
Supplement Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PSCST30 CARTRIDGE, LOT UNKNOWN
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-