Model Number MMT-751LNAS |
Device Problem
Sticking (1597)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
|
Event Description
|
Customer reported via phone call that the insulin pump had several issues.Customer¿s blood glucose was unknown at the time of incident.Customer reported that the insulin pump had a failed battery test alarm, rewind was loud, down button was sticking, and small bubbles on the screen.No troubleshooting was performed.Insulin pump is not expected to return for analysis.
|
|
Manufacturer Narrative
|
Insulin pump passed displacement test, self test, unexpected restart error test, off no power test and all operating currents tested within the spec range.No damage was found on lcd display found.Insulin pump received with intermittent button response at esc and act due to j2 connector unlocked on lcd board.
|
|
Manufacturer Narrative
|
Device passed displacement test, self test, a21 error test, off no power test and all operating currents tested within the specification range.No damage was found on lcd display found.Device received with intermittent button response at esc and act due to j2 connector unlocked on lcd board.(b)(4).
|
|