• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Moisture Damage (1405)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a power (moisture ingress) issue.It was noted that there was moisture behind the display.There was no allegation of an adverse event associated with this complaint.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
 
Additional Manufacturer Narrative
Follow-up #1: date of submission 24-jul-2018 device evaluation: the device has been returned and evaluated by product analysis on 09-jul-2018 with the following findings: during investigation, moisture was found behind the display lens.No damage was found to the returned battery cap.Unrelated to the original complaint, the battery compartment was cracked below the bumper pad.No moisture/corrosion was observed in the battery compartment.The returned battery cap fully attached to the pump.The pump had vibratory and auditory alarms with a blank display screen.A leak test failed due to a battery compartment leak.The pump cover was removed to find moisture ingress on the printed circuit board and internal components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7645078
Report Number2531779-2018-11916
Device Sequence Number13490809
Product Code OYC
UDI-Device Identifier10840406100334
UDI-Public10840406100334
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Was the Report Sent to FDA? Yes
Device Age19 MO
Initial Date Received by Manufacturer 06/17/2018
Supplement Date Received by Manufacturer06/17/2018
Initial Report FDA Received Date06/28/2018
Supplement Report FDA Received Date07/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age13 YR
-
-