| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a call service alarm (cs 064) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of submission (b)(6)2018 device evaluation: the device has been returned and evaluated by product analysis on (b)(6)2018 with the following findings: a review of the black box showed a call service 064 alarm in the pump history.The rewind, load and prime steps were performed successfully.The pump was exercised for 24 hours with no alarms occurring.Unrelated to the original complaint, corrosion was found in the battery compartment.Additionally, the battery compartment was cracked alongside of the pump.The pump leaked at the battery compartment.The pump was opened and moisture damage was found on the printed circuit board.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|