| Model Number |
MMT-722NAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
It was reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level at the time of the incident was unknown.The customer was disconnected during prime.The drive support cap was not damaged.It was advised to discontinue the use of the device and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Insulin pump passed displacement test, rewind test, basic occlusion test, prime/a33 test, excessive no delivery test and occlusion test.No compromised force sensor resistance alarm noted.
|
| |
|