Model Number 8813817005
Medical Device Problem Code Occlusion Within Device (1423)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Date of Event 05/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the procedure, the guide wire was not passing through the catheter.It was stated that povidone iodine was the cleaning agent used on the device.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to use, the guide wire was not passing through the catheter.It was stated that flushing was performed prior to use and it was successful.Another catheter was used and there was no issue with the guide wire.There was no patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted no abnormalities.Pmv performed functional testing, the guide wire was inserted and removed and reinserted into the tube without restriction.The catheter was submerged into a water bath.The end was clamped, and a syringe was used to inject air to observe leakage.No air bubbles were present.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
This event has been reassessed and found to be a non-mdr reportable complaint.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key7645347
Report Number3009211636-2018-00215
Device Sequence Number152438
Product Code FOZ
UDI-Device Identifier10884521004894
UDI-Public10884521004894
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K943349
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2021
Device Model Number8813817005
Device Catalogue Number8813817005
Device Lot Number1626700240
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Initial Date Received by Manufacturer 06/13/2018
Supplement Date Received by Manufacturer07/12/2018
09/07/2018
08/08/2018
08/08/2018
Initial Report FDA Received Date06/28/2018
Supplement Report FDA Received Date07/27/2018
09/27/2018
10/04/2018
10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-