| Model Number |
EEAORVIL21A |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic gastric banding procedure, while on gastrojejunostomy, it was difficult to remove the stapler from the cavity.The anvil did not tilt after firing and it was carefully worked out of the anastomoses.It took a few additional minutes to remove the anvil and this resolved the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the orvil cutting ring showed shallow traces of knife impression and the ring was received partially severed.Additionally, cross cutting staple line marks were also observed on the mylar cutting ring.Functional evaluation showed no abnormalities and the anvil tilted after firing.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.Replication of the observed knife blade damage may occur if the instrument is applied over an obstruction.The instructions for use manual for this product states: " make certain that the section of tissue to be stapled is free from any metal or similar structure; otherwise, the knife blade may not cut." if information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|