| Model Number |
DL-344-10-K |
| Medical Device Problem Codes |
Break (1069); High impedance (1291)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/31/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The explanted lead was not returned to neuropace for analysis, however review of the lead impedance and ecog data was performed and the data is consistent with a lead break.
|
| |
|
Event or Problem Description
|
|
In (b)(6) 2017, high impedance was noted on the port 1 left depth lead, indicating a possible lead break.Stimulation and detection were disabled on the lead by the neurosurgeon.The change was not anticipated to affect patient treatment since most events were noted to occur on the contralateral lead.In (b)(6) 2018, the lead was replaced.The lead was not returned to neuropace by the treating center so no additional investigation could be performed.
|
| |
|