(b)(4).A 13-month complaint / service history review for similar complaints was performed for the serial number (b)(4) from aware date of (b)(6) 2018, which included data from (b)(6) 2017 through (b)(6) 2018.There were no similar complaints identified during the searched period, which includes this event.The g8 operator's manual under chapter 6 troubleshooting is as follows: 6.4 abnormal chromatograms from patients with hemoglobin variants or unknown peaks not recognized by the analyzer are occasionally seen during routine testing.These patterns may indicate interferences or problems with the assay.Therefore, it is important to use caution when troubleshooting.Review all chromatograms to determine whether the results are valid.In most cases, results for the sa1c% are reportable.In some cases, the sa1c% may be invalid depending on the hemoglobinopathy present, the flow rate, and the condition of the column and reagent system.The most likely cause of the reported event is attributed to an unidentified hemoglobin variant hbae for the patient in question that interferes with the g8 high performance liquid chromatography (hplc) methodology.
|