MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G248
Medical Device Problem Codes Failure to Capture (1081); Failure to Sense (1559); Connection Problem (2900)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Date of Event 05/17/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event or Problem Description
Boston scientific received information that this system exhibited loss of sensing and capture on the atrial channel due to what appeared to be a setscrew problem.A revision procedure was performed and the associated competitive atrial lead was removed from service and replaced.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event or Problem Description
It was reported that dislodgement of the competitive atrial lead was the cause of the reported clinical observations on the initial report.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key7646655
Report Number2124215-2018-11367
Device Sequence Number16186981
Product Code NIK
UDI-Device Identifier00802526589300
UDI-Public00802526589300
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P010012/S436
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/09/2019
Device Model NumberG248
Device Catalogue NumberG248
Device Lot Number378180
Other Device ID NumberVIGILANT
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 05/17/2018
Supplement Date Received by Manufacturer02/25/2020
Initial Report FDA Received Date06/28/2018
Supplement Report FDA Received Date05/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
0184; 4671; 5076; 6948; 7740; G248; MISMATCH; T165; 0184; 4671; 5076; 6948; 7740; G248; MISMATCH; T165
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age71 YR
-
-