| Catalog Number |
TL60 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that during the resection of gastric cancer, after fired, found some of staples malformed with irregular shape.Another device was used to complete the surgery.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # m56236.Device evaluation: the analysis results showed that the tl60 device arrived with no apparent damage with a reload loaded on the device.The reload was returned void of staples.The device was tested for functionality with a test reload and it cycled, fired, and all the staples formed as intended.The device fired without any difficulties and the staples were noted to have the proper b-formed shape.The event described could not be confirmed as the device performed without any difficulties noted.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|
| |
|