• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); No Code Available (3191)
Date of Event 06/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Work order search: no similar complaint was reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -8.0 diopter, in the patient's right eye (od) on (b)(6) 2018.The lens was explanted on (b)(6) 2018 due to improper fit.Another lens was not implanted.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch will be submitted.
 
Additional Manufacturer Narrative
The lens was explanted on (b)(6) 2018 due to excessive vaulting, angle closure and elevated intraocular pressure.The incision was enlarged and suture was required.The reporter indicated the cause of the event was due to patient related factor.The lens was returned dry, in lens vial.Visual inspection found no visible damage to the lens.There was evidence of residue and debris on the lens.(b)(4).
 
Additional Manufacturer Narrative
One similar complaint type event were reported for units within the same lot.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7647578
Report Number2023826-2018-00984
Device Sequence Number11141060
Product Code MTA
UDI-Device Identifier00841542103251
UDI-Public00841542103251
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date08/31/2019
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/15/2018
Supplement Date Received by Manufacturer06/28/2018
09/18/2018
Initial Report FDA Received Date06/28/2018
Supplement Report FDA Received Date08/02/2018
09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL AND LOT NUMBER UNK; INJECTOR MODEL AND LOT NUMBER UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age27 YR
-
-