| Catalog Number |
HAP02 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Only event year known: 2018.Batch # unk.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Was the procedure converted to an open procedure? what part of the device had the hole? was it the blue iris seal that screws down around the surgeon¿s hand? was it the white part of the device that goes into the patient¿s body? did any piece of the device fall into the patient? if yes, how was the piece retrieved from the patient?.
|
| |
|
Event or Problem Description
|
|
It was reported that during a lower anterior bowel resection, while the surgeon was using device, he noticed gas leaking and realized there was a small hole in the device.There were no patient consequences, procedure was completed open.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch # p91w5r.Device analysis: the analysis results found that the hap02 device was received with the iris seal torn.However, it could not be determined what may have caused this condition.No conclusion could be reached as to what may have caused the reported incident.The reported complaint was confirmed.The batch history record was reviewed and identified a holes or tears defect related to the reported incident were found.When this occurs, our quality system documents the necessary actions to ensure final product quality.The final quality release criteria were met before this batch was released for distribution.In addition, no protocols or ncr related to the complaint, were found during the manufacturing process.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Upon review of the information provided, it was concluded that this event does not meet the fda defined criteria for a reportable event, and is being considered not reportable.Additional information: the surgeon normally uses a wound retractor device, so he tried this device, when it seemed to leak air, he continued the way he usually does with an open wound retractor.I noticed the hole on the blue iris seal, the one the tightens around the surgeons hand.There was no piece missing from the blue section, just a small hole that i suspect was caused by the surgeon or scrub sister while assembling.
|
| |
|