| Model Number |
H74939200091070 |
| Medical Device Problem Codes |
Bent (1059); Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage and migration occurred.The 75% stenosed target lesion was located in the mildly tortuous and severely calcified iliac artery.After pre-dilation, a 9x100x75mm epic¿ vascular stent was implanted for treatment.However, when physician tried to advance the sheath further while using dilator, it was noted that the edge of the stent was caught in the gap and got flattened.Upon pushing further, the stent shifted.The procedure was completed by implanting another stent inside the epic stent.No patient complications nor injuries were reported.The flow until the distal was also good.
|
| |
|