Unit passed the functional test, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and the delivery accuracy test.Insulin flow blocked alarm functions properly during the basic occlusion test, occlusion test and force sensor test.No unexpected insulin flow blocked alarms noted during testing.No active insulin anomaly noted.Unit communicated properly with the test blood glucose meter.The meter sent the test value of 4.8 mmol/l to the pump and was properly displayed on the pump screen.No pump/meter communication anomaly noted.The low reservoir alarm functions properly during testing.With 6 units remaining in the reservoir, programmed a 6.0 unit bolus delivery.After the pump completed the 6.0 unit bolus delivery, the pump triggered an alert, reservoir estimate at 0 unit.No low reservoir alarm anomaly or units left anomaly in the pump status screen noted.
|