Catalog Number 928852 |
Device Problem
Difficult to Advance (2920)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/06/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the plunger rod of a bd¿ insulin syringe with needle was ¿hard to move¿.There was no report of exposure, injury or medical intervention.
|
|
Manufacturer Narrative
|
Investigation summary: customer returned (11) loose 3/10cc, 12.7mm syringes.Customer states that the plunger is difficult to move.All returned syringes were tested and all were able to draw and expel properly without any observed defects.A review of the device history record was completed for batch# 7254995.All inspections and challenges were performed per the applicable operations qc specifications.There were three (3) notifications [(b)(4)] noted that did not pertain to the complaint.Severity: s_1__; occurrence: a complaint history check was performed and this is the 1st related complaint for plunger rod difficult to move on lot # 7254995 based on the samples / photo(s) received the investigation concluded: - unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure.Based on the above, no additional investigation and no capa is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Event Description
|
It was reported that the plunger rod of a bd¿ insulin syringe with needle was ¿hard to move¿.There was no report of exposure, injury or medical intervention.
|
|
Search Alerts/Recalls
|