Concomitant medical products: erbe electrosurgical generator.Investigation evaluation: an evaluation of the returned device confirmed the report on incorrect cutting wire orientation.During our laboratory analysis, the sphincterotome was advanced through a duodenoscope that was placed in a simulated biliary position.The duodenoscope has an accessory channel that is 4.2 mm in diameter (olympus model number tjf-160v).The catheter exited the endoscope with the cutting wire facing 9 o¿clock.The device was then bowed and the cutting wire was facing 10 o'clock (appropriate orientation is approximately 11:00 - 1:00 o'clock).The sphincterotome catheter was subjected to a close visual examination and twisting of the tubing was observed at the distal end.However, the device goes through several inspections in manufacturing, final quality control, and packaging departments prior to leaving the facility in an effort to ensure proper orientation.The inspection processes would have removed any products exhibiting twisting of the tubing prior to distribution.Therefore, it is unknown how or at what point the tubing became twisted at the distal end.Therefore, a discrepancy or anomaly that could have contributed to the reported event was not observed during our laboratory analysis of the returned product.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Incorrect cutting wire orientation can occur if the distal end of the catheter is shaped manually.This sphincterotome catheter is precurved and is provided with a precurved stylet in the distal tip of the catheter.This obviates the need for manual formation.The instructions for use contain the following comment: ¿note: do not apply manual pressure to tip or cutting wire of sphincterotome to influence orientation, as this may result in damage to device.¿ other factors that can contribute to improper cutting wire orientation include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all tri-tome pc triple lumen sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook tri-tome pc triple lumen sphincterotome.The device was advanced into 4.2 mm biopsy access endoscope accessory channel.The operator found that the cutting wire was pointing at 2 o'clock.There were multiple attempts to adjust the orientation of the cutting wire and they failed.At last, the operator replaced another of the same product to finish the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|