| Model Number |
N301 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient was not involved and there was not any impact or consequence to any patient.Device was not returned.Reserved samples from same lot tested and no failure detected.In addition reserved samples of current lot tested and continuity of wire and connection to the carbon disk was confirmed.Manufacturer was not able to reproduce the failure and is unable to confirm the complaint.
|
| |
|
Event or Problem Description
|
|
Customer reported that some of the neo leads n301 were not picking up any readings.
|
| |
|