Patient weight not available from the site.Initial troubleshooting, on the day following the event date, a medtronic representative went to the site to test the navigation system.It was reported that the image acquisitions were noted to not appear as transferred.Following a restart of the navigation system, the image acquisitions appeared.A medtronic representative went to the site to test the equipment.Testing revealed that the reported issue could not be replicated.The system then passed the system checkout and was found to be fully functional.
|
Medtronic received information that, while in a spinal fusion, the imaging system image acquisitions could not be received by the navigation system.It was reported that a second medtronic navigation system was brought into the operating room (or) to complete the procedure.It was noted that the navigation system displayed no error messages and that the communication link was green.There was no reported delay to the procedure due to this issue.There was no reported impact on patient outcome.No additional information was provided.
|