ABBOTT MANUFACTURING INC CLINICAL CHEMISTRY CARBON DIOXIDE Back to Search Results
Catalog Number 03L80-21
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A followup report will be submitted when the evaluation is complete.
 
Event or Problem Description
The account observed elevated architect carbon dioxide results ranging 33 to 35 mmol/l on 5 patient samples.The samples were repeated after recalibration with patient results in the account normal range of 20 to 31 mmol/l.No impact to patient management was reported.No specific patient information was provided.
 
Additional Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of field data, and a review of labeling.No adverse trend was identified for the customer's issue.No return patient sample was available.Historical performance of the reagent lot was evaluated using world wide data.The patient data was analyzed and within established control limits.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINICAL CHEMISTRY CARBON DIOXIDE
Common Device Name
CARBON DIOXIDE
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
MDR Report Key7649241
Report Number1628664-2018-00261
Device Sequence Number19300632
Product Code KHS
UDI-Device Identifier00380740005955
UDI-Public00380740005955
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K060295
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2018
Device Catalogue Number03L80-21
Device Lot Number51686UQ08
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/13/2018
Supplement Date Received by Manufacturer07/25/2018
Initial Report FDA Received Date06/28/2018
Supplement Report FDA Received Date08/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT C4000 LIST (B)(4); SERIAL (B)(4)
-
-