Model Number MMT-751NAL
Device Problem No Display/Image (1183)
Patient Problem Hyperglycemia (1905)
Event Date 05/28/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump had a flashing display.The customer¿s blood glucose level was unknown at the time of the incident.The customer was assisted with troubleshooting and the display did not return.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Insulin pump was received with blank display and constant tone due to moisture damage on the electronics assembly.Unable to perform all functional testing including the operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime/compromised force sensor system and excessive no delivery test or verify missing segment/partial display anomaly due to blank display.Insulin pump was received with moisture damage motor, vibrator, keypad and battery tube assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAL
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7649313
MDR Text Key112853015
Report Number3004209178-2018-86362
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503694
UDI-Public(01)00643169503694
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAL
Device Catalogue NumberMMT-751NAL
Device Lot NumberA4751NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age46 YR
Patient Weight157
-
-