Catalog Number 71342204
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 12/28/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to hip dislocation with bloodless reduction.
 
Additional Manufacturer Narrative
The associated polarcup insert and oxinium femoral head were not returned for evaluation.Our investigation including a review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.The devices were manufactured in 2016.A review of complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.A clinical evaluation noted that no clinical relevant documents were provided to conduct a thorough medical assessment.Thus, no medical assessment is warranted at this time.Without the return of the actual product involved, our investigation of this report is inconclusive.There is no information that would suggest the implanted devices failed to meet specifications.A relationship, if any, between the devices and the reported incident or adverse event could not be corroborated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXINIUM FEM HD 12/14 22 MM +4
Common Device Name
PRSTHSIS,HIP,SMI-CONSTRAIED,METAL/CERAMIC/POLYMR,CMNTDORNON-POROUS,UNCMNTD
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7649432
Report Number1020279-2018-01230
Device Sequence Number16189015
Product Code LZO
UDI-Device Identifier03596010488879
UDI-Public03596010488879
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K021673
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/30/2026
Device Catalogue Number71342204
Device Lot Number16DM00443
Initial Date Received by Manufacturer 06/01/2018
Supplement Date Received by Manufacturer06/01/2018
Initial Report FDA Received Date06/29/2018
Supplement Report FDA Received Date03/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age81 YR
-
-