COOK INC SPECTRUM TURBO-JECT ANTIBIOTIC POWER INJECTABLE PICC; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number G05676
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported spectrum turbo-ject antibiotic power injectable picc line had fractured and began leaking less than sixty days after placement.The patient required a general anesthetic to replace the line.The patient is currently at home on antibiotics.To date, additional information with regards to the patient, device and event are unknown.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information has previously been reported, is unknown or is unavailable.Correction investigation - evaluation a review of the complaint history, drawings, dimensional verification, device history record, documentation, instructions for use (ifu), manufacturing instructions, specifications, trends, quality control, and visual inspection of the returned device was conducted during the investigation.The device was returned with a secur-a-cath and microclave.Visual inspection found the extension tube to be slightly separated from the hub and a kink on the catheter at the distal end of the manifold.The device showed signs of normal use and cracks were not seen.A leak test confirmed the leak at the distal end of the hub where it connects to the extension tube.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product does not show any nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number and failure mode.Based on the information provided and the results of our investigation, a definitive root cause cannot be determined at this time, however appropriate measures have been initiated to address this failure mode.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRUM TURBO-JECT ANTIBIOTIC POWER INJECTABLE PICC
Common Device Name
FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7649665
Report Number1820334-2018-01858
Device Sequence Number11206254
Product Code FOZ
UDI-Device Identifier00827002056763
UDI-Public(01)00827002056763(17)190621(10)8387346
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/29/2018
Report Date (Section F) 06/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2019
Device Model NumberG05676
Device Catalogue NumberUPICS-3.0-CT-NT-ABRM-1110
Device Lot Number8387346
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Was the Report Sent to FDA? No
Event Location Hospital
Date Report to Manufacturer06/11/2018
Initial Date Received by Manufacturer 06/08/2018
Supplement Date Received by Manufacturer08/22/2018
Initial Report FDA Received Date06/29/2018
Supplement Report FDA Received Date09/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-