| Catalog Number |
7480111 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Type of procedure: extension on previous posterior lumber fusion levels implanted: t10-l5 it was reported that on (b)(6) 2018, intra-op, the tip on the screwdriver shaft broke off when the driver was engaged into the screw.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis result: visual and microscopic examination of the instrument identified shearing of the lead thread, consistent with incomplete engagement of the thread prior to loading.If information is provided in the future, a supplemental report will be issued.
|
| |
|