| Catalog Number |
309.521 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: it is unknown when device breakage occurred.Device is an instrument and is not implanted/explanted.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.A device history records review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that during decontamination procedure, four (4) devices were discovered to be not functioning.On a conical extraction screw for 1.5mm & 2.0mm cortex screws, the thread was damaged.Two instances of conical extraction screws for 3.5mm screws were broken.A spare reamer tube for hollow reamer was also broken.There was no patient or procedure involvement.This report is for a conical extraction screw for 3.5mm screws.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Dhr review was completed.Part: 309.521.Lot: l691951.Manufacturing site: bettlach.Release to warehouse date: 27.Feb.2018.The device history record shows this lot of 195 pieces was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and material criteria at the time of release with no issues documented during the manufacturing process.Measured hardness during manufacturing process was within specifications.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Customer quality conducted an investigation of the returned device.Investigation site: cq zuchwil.Selected flow: 5.Broken.Visual inspection: the received extraction screw shows that almost entire length of the threaded tip section has broken off.Dimensional inspection: because of the damages the complaint relevant dimensions cannot be checked to print specifications anymore.Drawing/specification review: the investigation has shown that the cause of complained malfunction is a post-manufacturing caused use related damage at the device, therefore no drawing/specification review is needed.Material review: the used material was stainless steel 1.4112 as required and the measured harness was with 668 - 678 hv10 within the specification of 510 - 720 hv10.The broken surface is homogenous what indicates material conformity as well.Summary: the received condition of the device is concordant with the complaint description and the complaint condition is confirmed.The manufacturing review shows that the production procedure was according to the specifications and there were no issues that would contribute to this complaint condition.Based on the provided information we are not able to determine the exact cause of this complaint.We only can assume that exceeding applied mechanical overload during the loosening of a blocked screw caused this breakage.Therefore, we can confirm the visible damages are not from any manufacturing non-conformity.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|