| Model Number |
VLOCL0013 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a trans abdominal pre peritoneal procedure, the loop broke.A new device was used to complete the case.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample noted one suture and one needle.The suture was received with the loop opened.Upon microscopic inspection, evidence of material transfer was observed at the weld site.Unfortunately, as no sealed samples were returned, a laced-through loop test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.A definitive root cause could not be determined with regard to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|