• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE THORACIC; SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE THORACIC; SCREW Back to Search Results
Model Number 24-016-07-74
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Reference exemption number e2017029.Multiple mdr reports were filed for this event, please see associated report: mdr 9610905-2018-00100.
 
Event or Problem Description
It was reported the patient's x-ray showed displaced plates and screws.
 
Additional Manufacturer Narrative
Two lot numbers were previously reported (33192743/33167051), their respective manufacture dates are 07-apr-2018 and 15-jun-2016.An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage falls within the design risk limits adhered to at klm.The review of the product device history records was performed on the lot numbers identified.In this investigation no anomalies were discovered.Due to no device being returned the failure root cause cannot be determined.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE THORACIC
Common Device Name
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM  78570
MDR Report Key7651579
Report Number9610905-2018-00101
Device Sequence Number11201162
Product Code HRS
UDI-Device Identifier00888118047731
UDI-Public(01)00888118047731
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K153482
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/06/2018
Report Date (Section F) 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number24-016-07-74
Device Lot Number33192743 & 33167051
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/06/2018
Event Location Hospital
Date Report to Manufacturer06/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2018
Supplement Date Received by Manufacturer10/12/2018
Initial Report FDA Received Date06/29/2018
Supplement Report FDA Received Date10/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age46 YR
-
-