• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BURR,5.5MM ABRADER,180MML DISPOSABLE (3; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BURR,5.5MM ABRADER,180MML DISPOSABLE (3; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200082
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/25/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a hip scope the burr used started spitting out blue stuff.No patient injury reported.
 
Additional Manufacturer Narrative
One 72200082 disposable 5.5mm 180mm abrader burr product returned.This was identified as demo device.The burr still spins semi freely due to the sluff chamber beginning to melt.This is where the "blue" fragments are coming from as the chamber melts and pulls fragments into the blade.This is a large scale cutting product which requires adequate irrigation to properly excise materials.Per ifu: "periodic irrigation of the blade is recommended to provide adequate cooling of the blade and to prevent accumulation of excised materials.Per ifu: "ensure that suction of 128 mmhg minimum is flowing while the instrument is running." root cause related to the manufacture of the device was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR,5.5MM ABRADER,180MML DISPOSABLE (3
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key7651726
Report Number1219602-2018-00850
Device Sequence Number10807132
Product Code HAB
UDI-Device Identifier03596010541529
UDI-Public(01)03596010541529(17)200224(10)50681538
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/24/2020
Device Model Number72200082
Device Catalogue Number72200082
Device Lot Number50681538
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2018
Supplement Date Received by Manufacturer06/27/2018
Initial Report FDA Received Date06/29/2018
Supplement Report FDA Received Date07/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-