• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4002C0830
Medical Device Problem Codes Detachment Of Device Component (1104); Kinked (1339)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2018-01301.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the splenic artery (spa)/ common hepatic artery (cha) using pod6s and a penumbra coil 400 (pc400).During the procedure, while attempting to insert a pod6 into the hub of a non-penumbra microcatheter, the physician experienced resistance at the distal tip of the pod6 introducer sheath; therefore, the physician removed the pod6 and soaked it in a saline trey.The physician then attempted again to advance the pod6 into the hub of the microcatheter; however, the pod6 became stuck at the same position in its introducer sheath.Therefore, it was removed.While attempting to advance a pc400 through the microcatheter, the pc400 pusher became kinked and subsequently, the pc400 unintentionally detached in the microcatheter.It was reported that the proximal end of the pc400 was approximately five centimeters into the microcatheter.Therefore, the physician removed the microcatheter containing the pc400 pusher assembly, the detached coil and removed the coil.The procedure was completed using the same microcatheter, four pod coils, one penumbra smart coil (smart coil) and four others coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked approximately 75.0 cm and fractured approximately 76.0 cm from the proximal end.The embolization coil was detached from its pusher assembly.The pc400 pusher assembly was fractured and, therefore, the pc400 could not be functionally tested.Conclusion: evaluation of the returned pod6 revealed that the pusher assembly was kinked.If the device is forcefully advanced against resistance, damage such as a kink may occur.During the functional test, resistance was encountered while attempting to retract the kinked pusher assembly through its introducer sheath due to the pusher assembly kinks and the pod6 could not be retracted any further.Further evaluation of the returned pod6 revealed that the introducer sheath had coagulated blood inside.This typically occurs if continuous flush is not used during the procedure.The kink in the pusher assembly and the coagulated blood inside the introducer sheath prevented the pod6 from being functionally tested during evaluation.Evaluation of the returned pc400 revealed that the pusher assembly was fractured.If the device is forcefully advanced against resistance, damage such as a kink may occur.Subsequently, if the kinked device is further manipulated the device may become fractured.Further evaluation of the returned pc400 revealed that the embolization coil was detached from its pusher assembly.If the two fractured segments of the pusher assembly are separated the pull wire will likely retract out of the ddt, detaching the embolization coil.Evaluation of the non-penumbra microcatheter revealed that the device was kinked in multiple locations.A demonstration coil was attempted to be advanced through the microcatheter; however, resistance was encountered and the coil could not be advanced any further.The complaint states that the same microcatheter was used to complete the procedure indicating these damages likely occurred post procedure.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-01301.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7651785
Report Number3005168196-2018-01302
Device Sequence Number14062382
Product Code HCG
UDI-Device Identifier00814548010564
UDI-Public00814548010564
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/05/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/09/2025
Device Catalogue Number4002C0830
Device Lot NumberF79970
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/05/2018
Supplement Date Received by Manufacturer11/19/2018
Initial Report FDA Received Date06/29/2018
Supplement Report FDA Received Date02/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Date Report Sent to FDA01/01/2005
-
-