| Model Number |
35164 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample is reported to be available, but has not yet been received by the manufacturer.The device history record for lot cm7361002 was reviewed and the product was produced according to product specifications.All information reasonably known as of 26 jun 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the pilot line on the microcuff came away from the tube during extubation.It remained in the mouth when the tube was removed.The pilot line was removed intact, but a direct laryngoscopy was performed to ensure there was nothing left in the airway.No further information has been provided at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
All information reasonably known as of 02 jul 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a halyard health product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
Per additional information received 25 jun 2018, the patient has been discharged from the hospital.
|
| |
|
Additional Manufacturer Narrative
|
|
One sample was received for evaluation.The sample was received with the inflation line detached from the tubing.The area of separation was examined under magnification.The endotracheal tube skive which receives the inflation line had a jagged fragment of the inflation line inside.The inflation line was severed flush with the skive opening.Under uv light, both the inflation line and the skive fluoresce, indicating that the bonding agent containing adhesive brightener was present.No root cause could be determined.All information reasonably known as of 04 feb 2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|