If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement the complaint device is not being returned, therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
|
The sales rep reported that during an acl procedure that two of his biointrafix tapered screws and large sheaths broke upon insertion.The sales rep stated that the surgeon had predrilled the hole and did not tap the devices in.The sales rep reported that the bone quality was very hard and he believed the surgeon may have been off axis.The surgeon completely removed both of the screws and sheaths and completed the procedure with a third similar set of devices with no patient consequences or delay.The devices were already discarded.Please see also medwatch report 1221934-2018-51030, 1221934-2018-51031, and 1221934-2018-51034.
|