| Catalog Number |
9733449 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.No parts have been received by the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of procedure.It was reported that a representative discovered two damaged straight suctions at the site.
|
| |
|
Additional Manufacturer Narrative
|
|
The straight suction was returned to the manufacturer for analysis.Analysis found that the instrument would not verify when attached to a known good tracker.Divot error 2.9 to 3.1 mm.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
| |
|