Model Number ORCHESTRA PLUS |
Device Problems
Electrical /Electronic Property Problem (1198); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/13/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that the subject programmer failed leakage current tests because it outputting too much energy (reportedly: "2 ohms which is 4 times the limit").Tests made on the power cable and the programmer reportedly determined the cable is fine, the issue is the programmer.
|
|
Event Description
|
It was reported that the subject programmer failed leakage current tests because it outputting too much energy (reportedly : "2 ohms which is 4 times the limit").Tests made on the power cable and the programmer reportedly determined the cable is fine, the issue is the programmer.
|
|