MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Device Problems Over-Sensing (1438); High Readings (2459); Device Displays Incorrect Message (2591)
Patient Problem Hyperglycemia (1905)
Event Date 06/01/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 60 mg/dl and the sensor glucose were 90 mg/dl at the time of the incident.Insulin delivery was suspended due to sensor glucose values.The customer was advised issue is most likely due to sensor not situated properly preventing the sensor from properly tracking changes in glucose.Customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.Suspend on low limit in sensor settings is 80.Customer was experiencing high blood glucose reading of 410.Customer did not troubleshoot their high blood glucose reading.Products will not return for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7653378
MDR Text Key112838037
Report Number3004209178-2018-86716
Device Sequence Number0
Product Code OZO
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/18/2018
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG2EJ1B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age71 YR
-
-