Model Number MMT-751LNAS
Device Problems False Reading From Device Non-Compliance (1228); Device Displays Incorrect Message (2591)
Patient Problem Hyperglycemia (1905)
Event Date 06/01/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned. No conclusion can be drawn at this time. We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm. The customer¿s blood glucose was 60 mg/dl and the sensor glucose were 90 mg/dl at the time of the incident. Insulin delivery was suspended due to sensor glucose values. The customer was advised issue is most likely due to sensor not situated properly preventing the sensor from properly tracking changes in glucose. Customer was informed that their blood glucose and sensor glucose levels were not in acceptable range. Suspend on low limit in sensor settings is 80. Customer was experiencing high blood glucose reading of 410. Customer did not troubleshoot their high blood glucose reading. Products will not return for analysis.
 
Manufacturer Narrative
The concomitant product has been added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name530G INSULIN PUMP MMT-751LNAS
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7653385
MDR Text Key112838878
Report Number3004209178-2018-86718
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model NumberMMT-751LNAS
Device Catalogue NumberMMT-751LNAS
Device Lot NumberA5751LNASJ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/03/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 06/29/2018 Patient Sequence Number: 1
-
-