• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754LWWB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754LWWB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-754LWWB
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/31/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
It was reported that their insulin pump received the drive support cap was stick out.The customer¿s blood glucose level was 238mg/dl.Troubleshooting was performed.The insulin pump will returned for analysis.
 
Manufacturer Narrative
Drive support disk was inspected and no anomaly was noted.Device received with cracked/detached end cap.Unable to perform operating currents, self-test, unexpected restart error test, rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test due to cracked/detached end cap.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754LWWB
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7653584
MDR Text Key112887452
Report Number3004209178-2018-86811
Device Sequence Number0
Product Code OYC
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-754LWWB
Device Catalogue NumberMMT-754LWWB
Device Lot NumberA2754LWWBJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Date Manufacturer Received07/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-