The affected device was returned and evaluated.A visual inspection of the 7mm starter drill confirmed the complaint that the tip of the drill has snapped off.The severed piece was not shipped with instrument.We recommend that all reusable instruments be routinely inspected for wear and damage so they can be replaced as necessary.The complaint history revealed no additional complaints for the associated batch.Additionally, a review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Insufficient clinically relevant documentation was provided to perform a thorough medical assessment on the root cause of the instrument breakage; however, overload fractures can occur if the mechanical loads applied to the re-use compression screw drill exceed the strength of the material.The 17-4 ph ss alloy is not approved/intended for implantation.The patient impact beyond the 30+ minutes delay and pain during the attempted extraction procedure itself cannot be determined, as the non-implantable foreign body/drill tip of unknown size reportedly remains within the patient in an unknown location.Although it was communicated that the surgeon did not anticipate any further patient impact, the patient's current status remains unknown, as well as if micro-motion is likely to occur.Based on this investigation, the need for corrective action is not indicated.No additional actions are being taken at this time; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.Credit has been issued for this device.
|