Model Number 383400 |
Device Problem
Decreased Pump Speed (1500)
|
Patient Problem
No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The customer reported the enteral feeding pump is running slow.There was no patient injury.
|
|
Manufacturer Narrative
|
An evaluation of the kangaroo pump was performed for the reported condition of the enteral feeding pump is running slow and stops in the middle of feeds.The unit was triaged and the reported issue could not be confirmed at this time.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The customer reported the enteral feeding pump is running slow and stops in the middle of feeds.There was no patient injury.On (b)(6) 2018, the customer stated that the pump underfed.The volume to be infused was set to 600 and the rate was set to 100.The actual volume delivered was close to 500.There was no harm to the patient.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|