The alleged broken ckp-4500 is not expected to be returned for evaluation and review.This complaint of broken device is unable to be verified and a root cause cannot be determined.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 27 complaints, regarding 29 devices, for this device family and failure mode.During this same time frame 76,496 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.0004.Per the instructions for use, the user is advised the following: warnings: surgeon must choose proper implant size based on specific procedure and patient history.Precautions: ensure proper selection of bone punch according to anchor size selection.Avoid lateral loading while inserting the poplok knotless suture anchor.Maintain proper alignment during insertion of the anchor and disengagement of the driver.Proper orientation and alignment of instruments is important during implantation of the poplok knotless suture anchor to minimize possible breakage of the anchor.Breakage of the poplok knotless suture anchor is possible if: the bone punch is not inserted to the proper depth.The poplok knotless suture anchor is not properly aligned with the pilot hole.The poplok knotless suture anchor inserter is used for prying.This issue will continue to be monitored through the complaint system to assure patient safety.
|
|
This distributor reported that the ckp-4500, 4.5mm poplok suture anchor, anchor broke into 3 pieces after insertion into the bone during a rotator cuff repair on (b)(6) 2018 causing a 2-minute delay in the procedure.All pieces were removed using a grasper.No medical intervention or extended hospital stay was required for the patient; as the patient was reported having no injuries.The reporter states that the patient was not "affected" by the event.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|