| Catalog Number |
47369802 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the seal was found open on a bd¿ pen needle sterile single use breaching sterility.There was no report of exposure, serious injury or medical intervention.
|
| |
|
Event or Problem Description
|
|
It was reported that the seal was found open on a bd¿ pen needle sterile single use breaching sterility.There was no report of exposure, serious injury or medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: customer returned a photo of a 7mm, 23g pen needle.Customer states that the needle was unsealed.The attached photo was examined and exhibited and incomplete seal between the tear drop label and outer cover.A review of the device history record was completed for batch # 6111750 all inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Based on the samples / photo(s) received the investigation concluded: confirmed: bd was able to duplicate or confirm the customer¿s indicated failure investigation conclusion: unable to determine a root cause.
|
| |
|